Simple optical readout for ethanol-chlorobenzene dosimetry system (CROSBI ID 467794)
Prilog sa skupa u zborniku | sažetak izlaganja sa skupa | međunarodna recenzija
Podaci o odgovornosti
Ilijaš, Boris ; Ražem, Dušan
engleski
Simple optical readout for ethanol-chlorobenzene dosimetry system
Optical readout of the ethanol-chlorobenzene (ECB or Dvornik dosimetry system) is based on the development of coloured secondary complex of ferric thiocyanate which has a maximum absorption at 485 nm. The applicability of a rugged, hand-held, battery powered filter colorimeter operating at 480 nm has been investigated as a reader for this purpose. This simple reader performs very well within absorbance one displaying an excellent linearity of absorbance with the concentration of Cl- ions. It is shown that by choosing the appropriate dilution factor when preparing the secondary complex solution the entire useful dose range of the dosimeter up to 2 MGy can be covered. The applicability of the same reader to some other liquid chemical dosimeters is also discussed.
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
Podaci o prilogu
1998.
objavljeno
Podaci o matičnoj publikaciji
Podaci o skupu
International Symposium on Techniques for High-Dose Dosimetry in Industry, Agriculture and Medicine
predavanje
02.11.1998-05.11.1998
Beč, Austrija