Determination of Stevioside by Square-wave Voltammetry (CROSBI ID 567002)
Prilog sa skupa u zborniku | sažetak izlaganja sa skupa | međunarodna recenzija
Podaci o odgovornosti
Komorsky-Lovrić, Šebojka ; Novak, Ivana ; Novak, Bruno
engleski
Determination of Stevioside by Square-wave Voltammetry
Steviol glycosides are electroactive compounds and can be detected by mercury electrode. The maxima of square-wave voltammograms of steviol and stevioside appear at -1.3 V vs Ag/AgCl in 1 M NaCl at approximately pH 11. The responses are quasireversible and probably caused by electro-reduction of carboxyl, or ester groups, respectively. The net peak current is linearly proportional to steviol concentration between 1 x 10-6 mol/L and 3.5 x 10-5 mol/L. This calibration plot can be used for the determination of the concentration of stevioside in extracts of stevia (Stevia rebaudiana Bertoni) leaves. The proposed method is applicable for routine analysis of stevioside in commercial products.
Stevioside; Square-wave Voltammetry
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
Podaci o prilogu
204-204.
2010.
nije evidentirano
objavljeno
Podaci o matičnoj publikaciji
Book of Abstracts
B. Tribollet, C. Amatore
Nica: ISE
Podaci o skupu
61 st Annual Meeting of the International Society of Electrochemistry
poster
26.09.2010-01.10.2010
Nica, Francuska