Proton dosimetry in presence of low energy X rays using TLD-700 (CROSBI ID 475984)
Prilog sa skupa u zborniku | izvorni znanstveni rad
Podaci o odgovornosti
Miljanić, Saveta ; Ranogajec-Komor, Maria ; Blagus, Saša ; Miljanić, Đuro ; Osvay, Margit
engleski
Proton dosimetry in presence of low energy X rays using TLD-700
LET dependence of high sensitive Al2O3:C and CaF2:Mn thermoluminescent (TL) dosemeters were investigated for proton dosimetry purposes. The experiments were performed with 137-Cs gamma rays and 2, 4 and 8 MeV protons using a Tandem van de Graaff accelerator. Experimental set up developed for the irradiation with protons is described. The irradiation dose was about 2 mGy for gammas and in the range 1-40 mGy for proton irradiation, respectively. The possibility of using these dosemeters for accelerator dosimetry purpose is discussed.
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
Podaci o prilogu
1999.
objavljeno
Podaci o matičnoj publikaciji
Deme, Sandor
Budimpešta: Eötvös Loránd University
Podaci o skupu
IRPA Regional Congress on Radiation Protection in Central Europe
poster
23.08.1999-27.08.1999
Budimpešta, Mađarska