Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi

Extensive asymptomatic pulmonary air embolism during CRT-implantation : Case report (CROSBI ID 213638)

Prilog u časopisu | stručni rad

Stein, A. ; Vražić, Hrvoje ; Semmler, V. ; Kolb, C. Extensive asymptomatic pulmonary air embolism during CRT-implantation : Case report // Acta cardiologica, 67 (2012), 5; 593-594. doi: 10.2143/AC.67.5.2174136

Podaci o odgovornosti

Stein, A. ; Vražić, Hrvoje ; Semmler, V. ; Kolb, C.

engleski

Extensive asymptomatic pulmonary air embolism during CRT-implantation : Case report

This report presents a case of pulmonary air embolism during implantation of a cardiac resynchronisation therapy system. For left ventricular pacing the thinnest available coronary sinus lead was used in association with a standard subclavian vein introducer sheath. It is hypothesised that the difference of size between the small lead diameter and the subclavian vein introducer sheath may have predisposed to air entry into the venous system.

CRT; complications; cardiovascular medicine