Use of a ruthenium(III), iron(II), and nickel(II) hexacyanometallate-modified graphite electrode with immobilized oxalate oxidase for the determination of urinary oxalate (CROSBI ID 130285)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Milardović, Stjepan ; Grabarić, Zorana ; Rumenjak, Vlatko ; Blau, Nenad ; Milošević, Danko Use of a ruthenium(III), iron(II), and nickel(II) hexacyanometallate-modified graphite electrode with immobilized oxalate oxidase for the determination of urinary oxalate // Journal of AOAC International, 84 (2001), 6; 1927-1933

Podaci o odgovornosti

Milardović, Stjepan ; Grabarić, Zorana ; Rumenjak, Vlatko ; Blau, Nenad ; Milošević, Danko

engleski

Use of a ruthenium(III), iron(II), and nickel(II) hexacyanometallate-modified graphite electrode with immobilized oxalate oxidase for the determination of urinary oxalate

This paper describes the performance of a biosensor with an Ru(III), Ni(II), and Fe(II) hexacyanometallate-modified graphite electrode and immobilized oxalate oxidase for the determination of urinary oxalate. The addition of ruthenium enhances the electrochemical reversibility and chemical stability of the electrocrystallized layer and improves the sensitivity of the biosensor. Hydrogen peroxide, produced by the enzyme-catalyzed oxidation of oxalate, was measured at -50 mV vs an Hg\Hg2Cl2\3M KCl electrode in a solution of pH 3.6 succinic buffer, 0.1M KCl, and 5.4mM ethylenediaminetetraacetic acid. The linear concentration range for the determination of oxalate was 0.18-280 muM. The recoveries of added oxalate (10-35 muM) from aqueous solution ranged from 99.5 to 101.7%, whereas from urine samples without oxalate (or with a concentration of oxalate below the detection limit) the recoveries of added oxalate ranged from 91.4 to 106.6%. The oxalate in 24 h urine samples, taken during their daily routine from 35 infants and children, was measured and found to range from 0.6 to 121.7 mg/L. There were no interferences from uric acid, acetylsalicylic acid, and urea in the concentration range investigated, but paracetamol and ascorbic acid did interfere. A good correlation (R-2 = 0.9242) was found between values obtained for oxalate in real urine samples by 2 laboratories, with the proposed biosensor and ion chromatography, respectively.

quartz-crystal microbalance; glassy-carbon electrode; chromium hexacyanoferrate; film electrodes; electrochemical-behavior; cupric hexacyanoferrate; thin-film; biosensor; oxidation; cyanide

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

84 (6)

2001.

1927-1933

objavljeno

1060-3271

Povezanost rada

Kemija

Indeksiranost