A study of the flexural strength and surface hardness of different materials and technologies for occlusal device fabrication (CROSBI ID 266319)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Prpić, Vladimir ; Slaćanin, Ivan ; Schauperl, Zdravko ; Ćatić, Amir ; Dulčić, Nikša ; Čimić, Samir A study of the flexural strength and surface hardness of different materials and technologies for occlusal device fabrication // The Journal of prosthetic dentistry, 121 (2019), 6; 955-959. doi: 10.1016/j.prosdent.2018.09.022

Podaci o odgovornosti

Prpić, Vladimir ; Slaćanin, Ivan ; Schauperl, Zdravko ; Ćatić, Amir ; Dulčić, Nikša ; Čimić, Samir

engleski

A study of the flexural strength and surface hardness of different materials and technologies for occlusal device fabrication

With the emergence of digital technologies, new materials have become available for occlusal devices. However, data are scarce about these different materials and technologies and their mechanical properties.The purpose of this in vitro study was to investigate the flexural strength and surface hardness of different materials using different technologies for occlusal device fabrication, with an emphasis on the digital technologies of computer-aided design and computer-aided manufacturing (CAD- CAM) and 3D printing. A total of 140 rectangular specimens were fabricated from two 3D-printed (VarseoWax Splint and Ortho Rigid), 2 CAD-CAM-produced (Ceramill Splintec and CopraDur), and 3 conventional autopolymerizing occlusal device materials (ProBase Cold, Resilit S, and Orthocryl) according to ISO 20795-1:2013. Flexural strength and surface hardness were determined for 10 specimens of each tested material using the 3- point bend test and the Brinell method. The data were analyzed using descriptive statistics and 1- way ANOVA with Bonferroni corrections (α=.05). Surface hardness values ranged from 28.5 ±2.5 MPa to 116.2 ±1.6 MPa. During flexural testing, neither the CopraDur nor the VarseoWax Splint specimens fractured during loading within the end limits of the penetrant's possible movement. Flexural strength values for other groups ranged from 75.0 ±12.0 MPa to 104.9 ±6.2 MPa. Statistical analysis determined significant differences among the tested materials for flexural strength and surface hardness. Mechanical properties among different occlusal device materials were significantly different. Acrylic resins were less flexible than polyamide and nonacrylic occlusal device materials for 3D printing but had higher and more consistent values of surface hardness. Clinicians should consider the different mechanical properties of the available materials when choosing occlusal device materials.

acrylic resin ; occlusal splint ; flexural strength ; surface hardness