Telemetry of the cochlear implant (CROSBI ID 501112)
Prilog sa skupa u zborniku | sažetak izlaganja sa skupa | međunarodna recenzija
Podaci o odgovornosti
Ries, Mihael ; Trotić, Robert ; Kekić, Branko ; Pegan, Boris ; Vagić, Davor ; Drviš, Petar ; Ajduk, Jakov ; Ćorluka, Tanja
engleski
Telemetry of the cochlear implant
Introduction: Telemetry of the cochlear implant has become the integral part of cochlear implantation procedure. A cochlear implant is now not only a stimulus-delivering device, but also a detector for electro-physiologic activity of the inner ear and auditory nerve... Patients and methods: In the period from June 1996 to the end of February 2004, 202 cochlear implantations were performed at our clinic with the same number of intraoperative telemetries. 82 telemetries were performed using the latest Cochlear software and interface available... Results: In the observed group we found that 9 (22%) had completely absent EESR. Three out of these 9 displayed unordinary NRT. One had to have his electrodes 13 to 19 measured manually, one had no response on electrodes 15-19 and the third showed no response on electrodes 1 and 19... Discussion: Altogether both EESR and ECAP measurements showed to be a reliable method of gathering useful data for future use with persons with cochlear implant...
cochlear implantation; telemetry; evoked potentials auditory
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
Podaci o prilogu
58-x.
2004.
objavljeno
Podaci o matičnoj publikaciji
Podaci o skupu
5 th European Congress of Oto-Rhino-Laryngology Head and Neck Surgery
predavanje
11.09.2004-16.09.2004
Rodos, Grčka