Accuracy Analysis of Dissolved Oxygen Measurement System with Cypress PSoC Configurable Mixed Signal Array (CROSBI ID 507136)
Prilog sa skupa u zborniku | izvorni znanstveni rad | međunarodna recenzija
Podaci o odgovornosti
Hrgetić, Mario ; Krois, Igor ; Cifrek, Mario
engleski
Accuracy Analysis of Dissolved Oxygen Measurement System with Cypress PSoC Configurable Mixed Signal Array
In this paper the accuracy analysis of the proposed dissolved oxygen measurement system based on the Clark's cell and the PSoC configurable mixed signal array chip is performed. In that process critical sources of errors are determined and associated errors are measured. The temperature, as the most influential parameter of oxygen measurements, is analyzed in greater detail. Monitoring and regulation of sample temperature is crucial for precision oxygen concentration measurements. In order to achieve optimal accuracy level, system construction implications are discussed and averaging algorithm and two point calibra-tion methods are applied.
programmable system on chip; PSoC; acccuracy analysis
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
Podaci o prilogu
1105-1110-x.
2005.
objavljeno
Podaci o matičnoj publikaciji
Proceedings of the IEEE International Symposium on Industrial Electronics 2005
Perić, Nedjeljko ; Petrović, Ivan ; Butković, Željko
Zagreb: Institute of Electrical and Electronics Engineers (IEEE)
Podaci o skupu
International Symposium on Industrial Electronics 2005, ISIE 2005
predavanje
20.06.2005-23.06.2005
Dubrovnik, Hrvatska