Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi

Analysis of the MRI used bird-cage resonator (CROSBI ID 32286)

Prilog u knjizi | izvorni znanstveni rad

Šimunić, Dina ; Wach, Paul Analysis of the MRI used bird-cage resonator // Electricity and Magnetism in Biology and Medicine / Bersani, Ferdinando (ur.). Bolonja: Plenum Press, 1999. str. 753-757-x

Podaci o odgovornosti

Šimunić, Dina ; Wach, Paul

engleski

Analysis of the MRI used bird-cage resonator

Calculation of RF electromagnetic energy deposition in the human body can be done by a numerical simulation or by calculating an analytical solution.The analysis performed here gives the resonant frequency calculation, followed by a calculation of a current distribution in the rods. This is of importance, because these currents give rise to the transverse RF magnetic field, which is necessary for nuclear magnetic resonance imaging. Therefore, the knowledge of the current distribution in the rods represents a basis for performing numerical simulations later on.

MRI, bird-cage resonator, analytical solution

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o prilogu

753-757-x.

objavljeno

Podaci o knjizi

Electricity and Magnetism in Biology and Medicine

Bersani, Ferdinando

Bolonja: Plenum Press

1999.

0306460416

Povezanost rada

Elektrotehnika