Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi

Expanding the functionality of HP4284A precision LCR meter through measurement system integration (CROSBI ID 529389)

Prilog sa skupa u zborniku | izvorni znanstveni rad | međunarodna recenzija

Smoljkić, Gabrijel ; Lacković, Igor Expanding the functionality of HP4284A precision LCR meter through measurement system integration // 13th International Conference on Electrical Bioimpedance and the 8th Conference on Electrical Impedance Tomography ICEBI 2007 ; IFMBE Proceedings ; Vol. 17 / Scharfetter, Hermann ; Merwa, Robert (ur.). Berlin : Heidelberg : New York: Springer, 2007. str. 232-235

Podaci o odgovornosti

Smoljkić, Gabrijel ; Lacković, Igor

engleski

Expanding the functionality of HP4284A precision LCR meter through measurement system integration

For many applications where frequency resolution is not critical, accurate impedance measurements from 20 Hz to 1 MHz can be obtained by HP4284A precision LCR meter. HP4284A is based on the auto-balancing bridge method which offers high accuracy over wide impedance range and relatively wide frequency coverage. In this paper we present a novel integrated architecture of the impedance measurement system. We have built the system around a PC, connected through Agilent 82357A USB/HPIB interface to the HP 4284A LCR impedance meter. Innovative and user friendly control of the instrument is accomplished by the computer application, which we have developed from scratch using the C# programming language and .NET 2.0 framework. The software application controls the instrument parameters and performs measurements through a defined measurement protocol. It enables both online and offline graphical presentation of measured data. To demonstrate the capabilities of the integrated impedance measuring system we present impedance measurements on a simple three-element RC network and on a physiological saline. Using our software application we precisely controlled the oscillator level, we acquired complex impedance data from the instrument and we simultaneously monitored voltage and current level on the object.

instrumentation; impedance measurement; HP 4284A; software; automatic parameter control

10.1007/978-3-540-73841-1_62 ; ISBN (Online) 978-3-540-73841-1

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o prilogu

232-235.

2007.

objavljeno

Podaci o matičnoj publikaciji

Scharfetter, Hermann ; Merwa, Robert

Berlin : Heidelberg : New York: Springer

978-3-540-73840-4

1727-1983

Podaci o skupu

International Conference on Electrical Bioimpedance and 8th Conference on Electrical Impedance Tomography (13 ; 2007)

predavanje

29.08.2007-02.09.2007

Graz, Austrija

Povezanost rada

Elektrotehnika