Development of a time-of-flight spectrometer at the Rudjer Boskovic Institute in Zagreb (CROSBI ID 135737)
Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija
Podaci o odgovornosti
Siketić, Zdravko ; Bogdanović Radović, Iva ; Jakšić, Milko
engleski
Development of a time-of-flight spectrometer at the Rudjer Boskovic Institute in Zagreb
A new TOF telescope has been constructed for thin film and surface analysis. The timing system consists of two electrostatic mirror 10 type detectors of Busch design. The detection efficiency of timing stations for very light ions was significantly improved using DLC (dia- 11 mond like carbon) foils coated with LiF instead of the conventionally used carbon foils. Ion energy is measured by a 300 mm2 ULTRA ion-implanted silicon detector. For the ERDA measurements with heavy and energetic ion beams, a time-of-flight (TOF) spectrometer is positioned at 37.5. Spectrometer can be easily moved to 120 backward angle for time-of-flight RBS analysis with low energy beam of light ions. Positioning and fine adjustments of sample orientation are performed with a motorized sample stage. The same spectrometer can be also installed at the ion microprobe scattering chamber for 3D elemental imaging.
Elastic recoil detection; Time-of-flight; Time resolution; Mass resolution; Depth resolution; Efficiency