Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi !

Analytical performance of Accu Chek Compact Plus blood glucometer (CROSBI ID 541177)

Prilog sa skupa u časopisu | sažetak izlaganja sa skupa | međunarodna recenzija

Šimundić, Ana-Maria ; Nikolac, Nora ; Kozar, Jelena ; Topić, Elizabeta Analytical performance of Accu Chek Compact Plus blood glucometer // Clinical chemistry and laboratory medicine / Siest, Gerard (ur.). 2008. str. A79-A79

Podaci o odgovornosti

Šimundić, Ana-Maria ; Nikolac, Nora ; Kozar, Jelena ; Topić, Elizabeta

engleski

Analytical performance of Accu Chek Compact Plus blood glucometer

Objectives: The aim of this study was to evaluate the analytical performance characteristics of the Accu Chek Compact Plus blood glucometer by comparing its accuracy and precision with our current laboratory reference analytical system – Olympus 2700 automatic analyser. Materials and methods: Original Olympus glucose reagent served as a reference method. The difference between glucose concentrations measured on reference and Accu Chek analytical platform was tested by paired t-test. Glucometer accuracy was evaluated using bias (%), Passing-Bablock regression, Bland-Altman and Error Grid Clark analysis. Intra-assay glucometer precision was examined by 10 consecutive measurements at two commercially available control samples and one random patient sample. Results: The average Accu Chek bias was -6.6% ; we found no significant deviation from linearity ; more than 95% of our values were between mean&plusmn ; ; 1.96 SD. The results of the Clarke Error Grid analysis were 100% in zone A. Conclusion: Accu Chek Compact Plus blood glucometer has satisfying accuracy and acceptable precision and may, therefore, be used interchangeably with our laboratory referent-system for glucose measurement.

blood glucometer; analytical performance

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o prilogu

A79-A79.

2008.

nije evidentirano

objavljeno

Podaci o matičnoj publikaciji

Clinical chemistry and laboratory medicine

Siest, Gerard

Berlin: Walter de Gruyter

1434-6621

Podaci o skupu

Swiss MedLab International Congress

poster

16.09.2008-19.09.2008

Montreux, Švicarska

Povezanost rada

Kliničke medicinske znanosti

Indeksiranost