High Resistance Compensation Method for Surface Potential Measurement (CROSBI ID 147123)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Kallay, Nikola ; Preočanin, Tajana ; Selmani, Atiđa ; Šupljika, Filip ; Leniček, Ivan High Resistance Compensation Method for Surface Potential Measurement // Croatica chemica acta, 82 (2009), 1; 323-327

Podaci o odgovornosti

Kallay, Nikola ; Preočanin, Tajana ; Selmani, Atiđa ; Šupljika, Filip ; Leniček, Ivan

engleski

High Resistance Compensation Method for Surface Potential Measurement

The High Resistance Compensation Method (HRCM) was developed to measure the electrode potentials of high resistance electrodes. It is essentially similar to the classical Poggendorff compensation method, but instead of an ampere-meter, the pH-meter was introduced as an indicator of complete compensation. The method was applied to measurements of the electrode potentials of a single crystal rutile (TiO2) electrode. Electrode potentials were converted to surface potentials, and the dependency of the surface potential on pH in aqueous medium was determined.

single crystal electrode; high resistance compensation; surface potential; rutile; Poggendorff