Note : Development of 9 A current source for precise resistance measurement method (CROSBI ID 187001)
Prilog u časopisu | kratko priopćenje | međunarodna recenzija
Podaci o odgovornosti
Štambuk, Igor ; Malarić, Roman
engleski
Note : Development of 9 A current source for precise resistance measurement method
In this Note, design of voltage controlled current source intended to be used in precise resistance measurement system in the range from 0.1 m to 10 Ohm is presented. The design specifications of current source include gross-tuning of current in the range from 0.1 mA to 9 A, low noise, low temperature coefficient, and short term stability better than 50 ppm. The realized current source has achieved better short term stability than comparable commercial devices.
current source ; noise ; resistance measurement
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
Podaci o izdanju
83 (106105)
2012.
106105-1-106105-3
objavljeno
0034-6748
10.1063/1.4759022