Time dependent accuracy of dental radiometers (CROSBI ID 193177)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Marović, Danijela ; Matić, Sanja ; Kelić, Katarina ; Klarić, Eva ; Rakić, Mario ; Tarle, Zrinka Time dependent accuracy of dental radiometers // Acta clinica Croatica, 52 (2013), 2; 173-180

Podaci o odgovornosti

Marović, Danijela ; Matić, Sanja ; Kelić, Katarina ; Klarić, Eva ; Rakić, Mario ; Tarle, Zrinka

engleski

Time dependent accuracy of dental radiometers

Inadequate intensity of the polymerization light source can compromise the quality and longevity of dental composite restorations. In order to maintain optimal strength of polymerization devices, regular control of polymerization units is necessary. The aim of this study was to compare the accuracy of two radiometers in measurement of light intensity of photopolymerization devices concerning the time point of the measurement. Light intensity measurements of 16 halogen and 8 LED curing lights were performed using three different devices, at the beginning as well as 10 and 40 seconds after the start of illumination. Two were handheld radiometers: Bluephase meter (BM) and Cure Rite (CR), while an integrating sphere (IS) represented the reference device. The data were statistically analyzed using Friedman’s test and Wilcoxon Signed Rank test (p<0.05). The values at the beginning and after 10 seconds measured by BM were significantly higher than measurements by IS, whereas CR showed higher values after 10 and 40 seconds. Both commercial radiometers tended to overestimate the light intensity of LED and halogen curing units when compared to the reference device. The time point of measurement influences the output value. The heating of radiometers was proposed as possible explanation for the inaccuracy.

dental curing lights ; radiometer ; LED curing units ; QTH curing lights