Analysis of the long-term stability of the Zener-based reference voltage standard (CROSBI ID 482326)
Prilog sa skupa u zborniku | izvorni znanstveni rad | međunarodna recenzija
Podaci o odgovornosti
Ilić, Damir ; Vujević, Dušan
engleski
Analysis of the long-term stability of the Zener-based reference voltage standard
The long-term stability of a DC Reference Standard Fluke 732A at 1 V, 1, 018 V and 10 V levels has been analysed according to the calibration data obtained in the Fluke, NIST and PTB laboratories over a period of more than fourteen years. The weighted least-squares fitting was used and the regression line coefficients, as well as those of the 3rd order polynomial, were calculated, by which the maintained voltages at all three outputs can be predicted for a moment of interest. This approach significantly reduces the uncertainties of the maintained voltages.
dc voltage standard; least-squares analysis; long-term drift
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
Podaci o prilogu
5-9-x.
2001.
objavljeno
Podaci o matičnoj publikaciji
Digest of the 18th Metrology Symposium
Ilić, Damir ; Boršić, Mladen ; Butorac, Josip
Cavtat: Hrvatsko mjeriteljsko društvo
Podaci o skupu
HMD 18th Metrology Symposium
predavanje
08.10.2001-10.10.2001
Cavtat, Hrvatska