Šestan, Danijel ; Jugović, Antun Evaluation of calibration bath with two precision thermometers // Knjiga radova jedanaeste CROLAB-ove konferencije KOMPETENTNOST LABORATORIJA 2015 / Lovorka Grgec Bermanec (ur.). Zagreb: CROLAB - Hrvatski laboratoriji, 2015. str. 10-15

Podaci o odgovornosti

Šestan, Danijel ; Jugović, Antun

engleski

Evaluation of calibration bath with two precision thermometers

The objective of this paper is to describe a specific procedure for evaluation of calibration baths used at Laboratory for Process Measurement (LPM) for high precision comparison calibrations of thermometers. Over time, improvements in technology lead to increased accuracies of transfer standards for temperature measurements, such as standard platinum resistance thermometers. Therefore, the performances of the calibration baths and furnaces used for comparison calibration are becoming a limiting factor in a process of temperature scale dissemination. Since the time variation of metrological characteristics of temperature regulated zones (gradients, stability, and repeatability) adds a significant contribution to the total calibration uncertainty, precise and accurate evaluation of those characteristics is of very high importance. The method described in this paper was used for evaluation of characteristics of liquid bath with the equalizing block. The bath was filled with water, and its characteristics were tested in a temperature range from 20 °C to 80 °C. The results of the tests indicate that combined expanded uncertainties of the bath lie in the range between 4.3 mK and 10.7 mK.

thermometry; calibration bath evaluation; thermometers; comparison calibration

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o prilogu

10-15.

2015.

objavljeno

Podaci o matičnoj publikaciji

Knjiga radova jedanaeste CROLAB-ove konferencije KOMPETENTNOST LABORATORIJA 2015

Lovorka Grgec Bermanec

Zagreb: CROLAB - Hrvatski laboratoriji

78-953-7329-18-1

Podaci o skupu

11th International Conference, Croatian Laboratories CROLAB

poster

21.10.2015-24.10.2015

Zadar, Hrvatska

Povezanost rada

Strojarstvo