A small current measurement system for gaseous micropattern detectors suitable for operation at high voltages (CROSBI ID 635161)

Prilog sa skupa u zborniku | sažetak izlaganja sa skupa

POLJAK, Nikola ; KOVAČIĆ, Marinko ; UTROBIČIĆ, Antonija ; PLANINIĆ, Mirko ; ERHARDT, Filip A small current measurement system for gaseous micropattern detectors suitable for operation at high voltages. 2015

Podaci o odgovornosti

POLJAK, Nikola ; KOVAČIĆ, Marinko ; UTROBIČIĆ, Antonija ; PLANINIĆ, Mirko ; ERHARDT, Filip

engleski

A small current measurement system for gaseous micropattern detectors suitable for operation at high voltages

We present a description of a high-resolution picoammeter developed in our laboratory. Tests of the prototype with a local Gas Electron Multiplier sheet setup demonstrate that it satisfies all the aforementioned requirements, including a current resolution of less than 15pA, a readout frequency of 1kHz, the ability of simultaneous readout of multiple channels (up to 16) and low battery power consumption.

Small current measurement; Floating picoammeter