Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi

A descriptive and broadly applicable model of therapeutic and stray absorbed dose from 6 MV to 25 MV photon beams (CROSBI ID 236504)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Schneider, Christopher ; Newhauser, Wayne D ; Jagetic, Lydia ; Schneider, Uwe ; Kaderka, Robert ; Miljanić, Saveta ; Knežević, Željka ; Stolarcyzk, Liliana ; Durante, Marco ; Harrison, Roger A descriptive and broadly applicable model of therapeutic and stray absorbed dose from 6 MV to 25 MV photon beams // Medical physics, 44 (2017), 7; 3805-3814. doi: 10.1002/mp.12286

Podaci o odgovornosti

Schneider, Christopher ; Newhauser, Wayne D ; Jagetic, Lydia ; Schneider, Uwe ; Kaderka, Robert ; Miljanić, Saveta ; Knežević, Željka ; Stolarcyzk, Liliana ; Durante, Marco ; Harrison, Roger

engleski

A descriptive and broadly applicable model of therapeutic and stray absorbed dose from 6 MV to 25 MV photon beams

Purpose was to develop a simple model of therapeutic and stray absorbed dose for a variety of treatment machines and techniques without relying on proprietary machine-specific parameters. Dosimetry measurements conducted in this study and from the literature were used to develop an analytical model of absorbed dose from a variety of treatment machines and techniques in the 6 MV to 25 MV interval. A modified one-dimensional gamma-index analysis was performed to evaluate dosimetric accuracy of the model on an independent dataset consisting of measured dose profiles from seven treatment units spanning four manufacturers. The average difference between the calculated and measured absorbed dose values was 9.9% for those datasets on which the model was trained. Additionally, These results indicate that the model can provide accurate calculations of both therapeutic and stray radiation dose from a wide variety of radiotherapy units and techniques. We have developed a simple analytical model of absorbed dose from external beam radiotherapy treatments in the 6 MV to 25 MV beam energy range. The model has been tested on measured data from multiple treatment machines and techniques and is broadly applicable to contemporary external beam radiation therapy.

out-of-field dose, absorbed dose, analytical model, external beam radiotherapy