Wavelength-dependent light transmittance in resin composites: practical implications for curing units with different emission spectra (CROSBI ID 263500)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Par, Matej ; Repušić, Igor ; Skenderović, Hrvoje ; Tarle, Zrinka Wavelength-dependent light transmittance in resin composites: practical implications for curing units with different emission spectra // Clinical oral investigations, 23 (2019), 12; 4399-4409. doi: 10.1007/s00784-019-02896-y

Podaci o odgovornosti

Par, Matej ; Repušić, Igor ; Skenderović, Hrvoje ; Tarle, Zrinka

engleski

Wavelength-dependent light transmittance in resin composites: practical implications for curing units with different emission spectra

Objectives To evaluate light transmittance as a function of wavelength for eight composite materials and compare the transmittance for blue light produced from two curing units with different emission spectra. Materials and methods Light transmittance through 2- and 4- mm-thick composite specimens was recorded in real time during 30 s of curing using a broad- spectrum (peaks at 405 and 450 nm) and a narrow-spectrum (peak at 441 nm) LED-curing unit. The spectral resolution of 0.25 nm and temporal resolution of 0.05 s resulted in a large amount of light transmittance data, which was averaged over particular spectral ranges, for the whole measurement period. Statistical analysis was performed using Welch ANOVAwith Games-Howell post hoc test, t test, and Pearson correlation analysis. The level of significance was 0.05 and n = 5 specimens per experimental group were prepared. Results Light transmittance varied as a function of wavelength and time, revealing significantly different patterns among the tested materials. Light transmittance for different parts of curing unit spectra increased in the following order of emission peaks (nm): 405 < 441 < 450. Of particular interest was the difference in transmittance between 441 and 450 nm, as these peaks are relevant for the photoactivation of camphorquinone-containing composites. A high variability in light transmittance among materials was identified, ranging from statistically similar values for both peaks up to a fourfold higher transmittance for the peak at 450 nm. Conclusion Each material showed a unique pattern of wavelength-dependent light transmittance, leading to highly materialdependent differences in blue light transmittance between two curing units. Clinical relevance Minor differences in blue light emission of contemporary narrow-peak curing units may have a significant effect on the amount of light which reaches the composite layer bottom.

Resin composites ; Light transmittance ; Visible light curing ; Wavelength ; Emission spectrum

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

23 (12)

2019.

4399-4409

objavljeno

1432-6981

1436-3771

10.1007/s00784-019-02896-y