Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi

Evaluation of a commercial real-time PCR assay for quantitation of Epstein-Barr virus DNA in different groups of patients (CROSBI ID 124479)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Kozić, Sanja ; Vince, Adriana ; Iščić Beš, Janja ; Đaković Rode, Oktavija ; Židovec Lepej, Snježana ; Poljak, Mario ; Božić, Michael ; Kessler, Harald H. Evaluation of a commercial real-time PCR assay for quantitation of Epstein-Barr virus DNA in different groups of patients // Journal of virological methods, 135 (2006), 2; 263-268-x

Podaci o odgovornosti

Kozić, Sanja ; Vince, Adriana ; Iščić Beš, Janja ; Đaković Rode, Oktavija ; Židovec Lepej, Snježana ; Poljak, Mario ; Božić, Michael ; Kessler, Harald H.

engleski

Evaluation of a commercial real-time PCR assay for quantitation of Epstein-Barr virus DNA in different groups of patients

The aim of this study was to evaluate the performance of a molecular assay for quantitation of Epstein-Barr (EBV) DNA based on real-time PCR, and to determine EBV DNA levels in EDTA whole blood samples derived from different groups of patients. Following a manual DNA extraction protocol, real-time PCR was performed using the LightCycler EBV Quantification Kit, which demonstrated sufficient accuracy and linearity. Coefficients of variation were found to be between 6 and 42% and 5 and 34% respectively, for inter-assay and intra-assay variations. In clinical specimens, EBV DNA was detected in all patients with acute EBV infection (n=34), in one of 25 audults with past EBV infection, in 16 out of 25 (64%) anti-HIV antibody positive persons, in 10 out of 25 (40%) solid organ transplant recipients, and in none of the 23 infants without the history of EBV infection. When EBV DNA levels in positive specimens were compared between different groups, statistically significant differences were not found. The LightCycler EBV quantification kit was found to be useful for determination of EBV DNA levels in EDTA whole blood.

real-time PCR; Epstein-Barr virus; quantitation; viral load; immunosupression

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

135 (2)

2006.

263-268-x

objavljeno

0166-0934

Povezanost rada

Temeljne medicinske znanosti, Kliničke medicinske znanosti, Biologija

Indeksiranost