Determination of lead in Croatian wines by graphite furnace atomic absorption spectrometry (CROSBI ID 170520)
Prilog u časopisu | kratko priopćenje | međunarodna recenzija
Podaci o odgovornosti
Tariba, Blanka ; Pizent, Alica ; Kljaković-Gašpić, Zorana
engleski
Determination of lead in Croatian wines by graphite furnace atomic absorption spectrometry
A method has been developed for direct determination of lead in wine by graphite furnace atomic absorption spectrometry (GFAAS) with Zeeman-effect background correction. The thermal behaviour of Pb during pyrolysis and atomisation stages was investigated without matrix modifier and in the presence of Pd(NO3)2, Pd(NO3)2 + Mg(NO3)2 x 6H2O, and NH4H2PO4 + Mg(NO3)2 x 6H2O as matrix modifiers. A simple 1:1 dilution of wine samples with Pd(NO3)2 as a matrix modifier proved optimal for accurate determination of Pb in wine. Mean recoveries were 106 % for red and 114 % for white wine, and the detection limit was 3 μg L-1. Within-run precision of measurements for red and white wine was 2.1 % and 1.8 %, respectively. The proposed method was applied for analysis of 23 Croatian wines. Median Pb concentrations were 33 μg L-1, range (16 to 49) μg L-1 in commercially available wines and 46 μg L-1, range (14 to 559) μg L-1 in home-made wines. There were no statistically significant differences (P<0.05) in Pb concentration between commercial and home-made wines or between red and white wines.
analytical validation; commercial wines; GFAAS; home-made wines
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
Podaci o izdanju
62 (1)
2011.
25-31
objavljeno
0004-1254
10.2478/10004-1254-62-2011-2073