Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi !

Influence of radiation Effect on The Measurement Relative Intensity in X-ray Emission Spectroscopy (CROSBI ID 483930)

Prilog sa skupa u zborniku | sažetak izlaganja sa skupa | međunarodna recenzija

Novosel-Radović, Vjera ; Radović, Nikol ; Balen, Milka Influence of radiation Effect on The Measurement Relative Intensity in X-ray Emission Spectroscopy // Book of Abstracts from 4th Mediterranean Basin Conference on Analytical Chemistry / Veber, Marijan (ur.). Ljubljana: Slovensko kemijsko društvo, 2002. str. 134-x

Podaci o odgovornosti

Novosel-Radović, Vjera ; Radović, Nikol ; Balen, Milka

engleski

Influence of radiation Effect on The Measurement Relative Intensity in X-ray Emission Spectroscopy

It has been observed that standard samples, such as iron ore, sinters, mixture for sinter and blast furnace slag pressed into briquet with boric acid (ratio 1:3) diminished their wear resistance after radiation. Also, the measured intensities of analytical lines CaK alpha, SiK alpha and FeK alpha showed the deviation concerning the accuracy and reproducibility of measurement. To explain those deviations an investigation was undertaken to find out how to the measured intensities were affected by radiation effect. The model samples were used. They were prepared with calcit, quartz, boric acid and scale (mixture of iron compounds, more than 90 % Fe3O4). The components were ground and separed in fractions of different grain size by sieve. Parts of separeted fractions were irradiated for 30, 60, 90 and 120 minutes in the Philips semi-automatic x-ray spectrometer. Irradiated samples of boric acid were examined by sieve analysis, and each fraction was again subjected to irradiated and sieve analysis. The samples of different grain size: calcit, quartz, scale and boric acid were homogenized and briqueted using a Spex Al-holder. Measurement of analytical lines of CaK alpha, SiK alpha, FeK alpha were made on a Philips sequential semi automatic x-ray spectrometer. Secondary electron micrographs of non - irradiated boric acid were taken on scanning electron microscope. The values of the measure relative intensities and sieve analysis were treated mathematicaly. The analysis showed that differences in the accuracy and reproducibility of measurement of the relative intensities of analytical lines of CaK alpha, SiK alpha, FeK alpha were due to the heterogeneity effect as results of radiation effect on the binder boric acid.

x-ray emission spektrometry; radiation effect; relative intensity

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o prilogu

134-x.

2002.

objavljeno

Podaci o matičnoj publikaciji

Veber, Marijan

Ljubljana: Slovensko kemijsko društvo

Podaci o skupu

4th mediterranean Basin Conference on Analytical Chemistry

poster

15.09.2002-20.09.2002

Portorož, Slovenija

Povezanost rada

Geodezija, Metalurgija